Description
- New and Improved touch-screen with full colour display. The new user friendly interface makes it a self-explanatory unit with the interactive learning on the go, including electrode placement, running EMG graph as well as many other helpful tools.
- Pre-set programmes for Incontinence treatment and neuro-muscular Rehabilitation.
- Designed for Physiotherapist in mind to be used as a portable tool. Suitable as a personal trainer for in-clinic or home use (under guidance).
- Single channel EMG (Biofeedback) combined with 1 channels of NMES (stimulation).
- Single channel ETS (EMG triggered stimulation) with single channel stimulation.
- Simple single-phase or multiphase operation: multiphase allows EMG/ETS/STIM to be combined in one programme.
- Multilingual LCD screen and voice prompts for Biofeedback.
- Manual and Automatic threshold control.
- EMG Biofeedback between stimulation (muscle activity and tiredness indication when using muscle stimulation).
- Stand alone device or used in conjunction with established multilingual software.
- Now with Bluetooth connectivity! The unit sends up to 10 metres of wireless data to PC software.
- For home users: Produces and prints (with the optional PC Software) the comprehensive patient progress reports on a day to day basis providing average readings (2 channels) on work/rest, peak, onset and muscle release times, work and rest deviations and the average mA current.
- Used as a learning tool for both patient and clinician with emphasis on improving treatment techniques in the use of EMG and Neuromuscular stimulation.
- 4 x AAA batteries {No mains required}.
- Model number: MYO120P. This product is available through selected distributors who are able to provide training and support.
Specifications
1. EMG
1.0 Single channel EMG
1.1 EMG Range: 0.2 to 2000 μV RMS (continuous)
1.2 Sensitivity: 0.l μV RMS
1.3 Accuracy: 4% of μV reading +/-0.3 μV at 200 Hz
1.4 Selectable Bandpass filter – 3db Bandwidth,
a. Wide: 18 Hz +/- 4 Hz to 370 Hz +/- 10% – Reading
below 235 microvolts
10 Hz +/-3 Hz to 370 Hz +/- 10% – Reading
above 235 microvolts
b. Narrow: 100 Hz +/- 5% to 370 Hz +/- 10%
1.5 Notch filter: 50 Hz (Canada 60Hz) – 33 dbs (0. 1% accuracy)
1.6 Common Mode Rejection Ratio: 130 dbs Minimum @ 50 Hz
1.7 Work / Rest periods: 2-99 seconds
1.8 Number of Trials: 2-99
2. NMES, TENS and ETS
2.1 Single channel Stimulator
2.2 Amplitude: 0-90 mA into 500 Ohm load – actual mA will tend to be less
than indicated due to Electrode impedance: at 1000 Ohms load (Electrodes
in poor condition) the maximum will be limited to 75 mA, at 1500 Ohms
load the maximum will be limited to 50 mA.
2.3 Type: Constant current, maximum output
voltage 70 Volts +5 / -10 Volts
2.4 Waveform: Symmetrical, rectangular,
bi-phasic with net zero DC current
2.5 Pulse width selection: 50-330 μS (10% accuracy)
2.6 Pulse rate selection: 2-100 Hz (5% accuracy)
2.7 Work / Rest periods: 2-99 seconds
2.8 Time 1 – 99 minutes
2.9 Ramp up time: 0.1 – 9.9 seconds
2.10 Preset and user programmable treatment Programs
2.11 Automatic output shut off with detection of open electrode above 10 mA
(+/- 2mA).
Battery:
Battery set: 4 x 1.5V, AAA battery.
Low battery indication at 5.4 volts +/- 0.2 volts, automatic shut off when voltage
drops below the low indication. Replace the batteries immediately!
The device goes off automatically when not in use (energy saving): for example
when in settings and no key pressed over 1 minute, when in stimulation
mode (home screen) and all channels are 0mA.
Expected average battery set life [of standard 850 mAh, alkaline]:
14-18 hours in STIM, 28 hours in EMG mode.
Expected service life:
5 years. Careful use and maintenance extends the life of the unit over the
service life limit.
Environmental Conditions for use:
+5 to +40 degrees Centigrade. 15-93% Humidity.
Environmental conditions for storage & transport:
-25 to +70 degrees Centigrade. 15-93% Humidity.
Physical:
Dimensions: Length 150 mm, Width 89 mm, Depth 35 mm.
Weight: MyoPlus Pro device: 140g (without batteries).